    The performance index of Jiangsu Tianling viscometer meets the requirements of metrological verification regulation
    Issue date:2022/03/08    Views:51    Classification:Technical

    1. The performance index of the instrument must meet the requirements of metrological verification regulation.


    The instruments in use shall be subject to periodic verification. If necessary (the instruments are used frequently or in a qualified critical state), intermediate self-examination shall be carried out to determine that their metrological performance is qualified, and the coefficient error is within the allowable range, otherwise accurate data cannot be obtained.


    2. Pay special attention to the temperature of the measured liquid.


    Many users ignore this point and think that the temperature difference doesn't matter at all. Our experiments have proved that when the temperature deviation is 0.5 ℃, the viscosity value deviation of some liquids exceeds 5%, and the temperature deviation has a great impact on the viscosity. When the temperature rises, the viscosity decreases. Therefore, pay special attention to keep the temperature of the measured liquid constant near the specified temperature point, and do not exceed 0.1 ℃ for accurate measurement.


    3. Selection of measuring container (outer cylinder).


    For the viscometer, carefully read the instrument manual, and different rotors (inner cylinder) match the corresponding outer cylinder, otherwise the measurement result will have a huge deviation.


    4. Select the rotor correctly or adjust the speed so that the indicated value is between 20-90 grids.


    This kind of instrument adopts the dial plus pointer method to read, and its stability and reading deviation are combined to have 0.5 grid. If the reading is too small, such as around 5 grid, the relative error caused is more than 10%. If the appropriate rotor or speed is selected to make the reading at 50 grid, the relative error can be reduced to 1%. If the indicated value is above 90 grids, the torque generated by the hairspring is too large, which is easy to creep and damage the hairspring. Therefore, the rotor and speed must be correctly selected.


    5. Frequency correction.


    The nominal frequency of domestic instruments is 50Hz, and the current power supply frequency in China is also 50Hz. We use a frequency meter to test the variability of less than 0.5%, so the general measurement does not need frequency correction. However, for some instruments in Japan, Europe and the United States, the nominal frequency is 60Hz, so the frequency must be corrected, otherwise 20% error will occur. The correction formula is: actual viscosity = indicated viscosity × Nominal frequency ÷ actual frequency.


    6. The depth of the rotor immersed in the liquid and the influence of bubbles.


    The rotary viscometer has strict requirements on the depth of the rotor immersed in the liquid, and must be made according to the requirements * of the manual. There are often bubbles in the process of rotor immersion in liquid, and most of them will float up and disappear after the rotor rotates for a period of time. Sometimes the bubbles attached to the lower part of the rotor cannot be eliminated, and the existence of bubbles will bring great deviation to the measurement data, so dipping the rotor slowly and obliquely is an effective method.


    7. Rotor cleaning.


    The rotor (including the outer cylinder) used for measurement should be clean and free of dirt. Generally, it should be cleaned in time after measurement, especially after measuring paint and adhesive. Pay attention to the cleaning method. Soak it in a suitable organic solvent. Never scrape it hard with a metal knife, because serious scratches on the rotor surface will lead to the deviation of the measurement results.


    National free consultation hotline:400-928-3898

    Mailbox:jstenlin17@163.com

    Address: No. 4, Zhonglian Road, Baocai Industrial Park,
    Yandu District, Yancheng City.

    CopyRight ? 2022 Jiangsu Tianling Instrument Co., Ltd. copyright   Su ICP Bei 17018543-1
    主站蜘蛛池模板: 亚洲粉嫩av一区二区黑人 | 敦煌市| 手机色在线| 青青青国产免A在线观看| 国产精品综合色区av| 国产精品自线在线播放| 中文字幕亚洲精品人妻| 精品午夜一区二区三区| 炉霍县| 阿克陶县| 午夜福利视频男同女同| 亚洲AV无码成人精品区日韩密殿| 白白色发布在线播放国产| 蜜桃av一区二区高潮久久精品| 亚洲乱码一区AV春药高潮| 夫妻一起自拍内射小视频| 熟妇与小伙子露脸对白| 上栗县| 南昌市| 长治市| 司法| av天堂线上| 江源县| 久久精品成人亚洲另类欧美 | 久久国产亚洲一区二区三区| 宁波市| 平邑县| 塘沽区| 人妻无码ΑV中文字幕久久琪琪布| 车险| av福利免费在线观看网站| 久操加勒比视频在线观看| 久久精品国产热久久精品国产亚洲| 亚洲免费视频一区二区三区| 久久久久AV成人无码网站| 河曲县| 亚洲av色香蕉一区二区蜜桃| 亚洲一区二区三区在线观看播放| 欧美丝袜激情办公室在线观看| 亚洲中文字幕日产喷水| 欧美精品v欧洲高清|